基于受激参量下转换的红外标准传递辐射计定标技术

盛文阳 夏茂鹏 李健军 翟文超 郑小兵

盛文阳, 夏茂鹏, 李健军, 翟文超, 郑小兵. 基于受激参量下转换的红外标准传递辐射计定标技术[J]. 红外与激光工程, 2019, 48(12): 1204001-1204001(7). doi: 10.3788/IRLA201948.1204001
引用本文: 盛文阳, 夏茂鹏, 李健军, 翟文超, 郑小兵. 基于受激参量下转换的红外标准传递辐射计定标技术[J]. 红外与激光工程, 2019, 48(12): 1204001-1204001(7). doi: 10.3788/IRLA201948.1204001
Sheng Wenyang, Xia Maopeng, Li Jianjun, Zhai Wenchao, Zheng Xiaobing. Calibration technology of infrared standard transfer radiometer based on stimulated parametric down-conversion[J]. Infrared and Laser Engineering, 2019, 48(12): 1204001-1204001(7). doi: 10.3788/IRLA201948.1204001
Citation: Sheng Wenyang, Xia Maopeng, Li Jianjun, Zhai Wenchao, Zheng Xiaobing. Calibration technology of infrared standard transfer radiometer based on stimulated parametric down-conversion[J]. Infrared and Laser Engineering, 2019, 48(12): 1204001-1204001(7). doi: 10.3788/IRLA201948.1204001

基于受激参量下转换的红外标准传递辐射计定标技术

doi: 10.3788/IRLA201948.1204001
基金项目: 

中国科学院实验室基金(CXJJ-17S003);国家自然科学基金(41805019)

详细信息
    作者简介:

    盛文阳(1991-),女,博士生,主要从事遥感器高精度定标方面的研究。Email:shengwy@mail.ustc.edu.cn

  • 中图分类号: O432.1

Calibration technology of infrared standard transfer radiometer based on stimulated parametric down-conversion

  • 摘要: 基于受激参量下转换过程产生光子的相关特性,构建了一套无需制冷且只利用可见光探测器即可实现红外标准传递辐射测量的系统。首先详细描述了该红外标准传递辐射计系统的工作原理,接着利用标准黑体拟合获取了该系统的定标系数,由此得到红外标准传递辐射计温度与待测黑体响应值的关系,最后通过与水浴黑体在相同温度点的响应值所反演获取的辐射温度作比较,进一步评估和验证了该红外标准传递辐射计系统的不确定度。实验结果表明:该红外标准传递辐射计测量水浴黑体辐亮度的联合不确定度为1.64%,表明该红外标准传递辐射计可作为标准黑体到用户传感器的传递标准。
  • [1] Xu Jun, Meng Binghuan, Zheng Xiaobing, et al. Calibration of common temperature blackbody based on thermal-infrared standard radiometer[J]. Infrared and Laser Engineering, 2013, 43(3):716-721. (in Chinese)
    [2] Zhang Zhao, Wan Zhi, Sun Jingxu, et al. Development and calibration of dual-waveband infrared radiometer for jones calibration method[J]. Acta Optica Sinica, 2016, 36(6):0623002. (in Chinese)
    [3] Yuan Xiaochun, Yang Zhixiong, Yu Chunchao, et al. Study of radiation calibration for LWIR hyper spectral imager spectrometer[J]. Infrared Technology, 2015, 37(5):431-434. (in Chinese)
    [4] Gong Luyu, Hao Xiaopeng, Sun Jianping, et al. Development of H500 type high precision vacuum blackbody used as a calibration standard for infrared remote sensing[J]. Acta Metrologica Sinica, 2017, 38(2):129-134. (in Chinese)
    [5] Li Yingwen, Chen Fusheng, Yang Changcheng, et al. Introduction to inflrared test standard transfer[J]. Infrared and Laser Engineering, 2007, 36(5):119-123. (in Chinese)
    [6] Rice J P, Johnson B C. The NIST EOS thermal infrared transfer radiometer[J]. Metrologia, 2003, 35(4):505-509.
    [7] Monte C, Gutschwager B, Morozova J S P, et al. Radiation thermometry and emissivity measurements under Vacuum at the PTB[J]. International Journal of Thermophysics, 2009, 30(1):203-219.
    [8] Xu Jun. Research on calibration of ambient temperature blackbodies based on a thermal-infrared standard Radiometer[D]. Hefei:Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, 2013. (in Chinese)
    [9] Xie Chenyu, Li Jianjun, Pang Weiwei, et al. Reseach on calibration method of double-channel thermal-infrared standard radiometer[J]. Infrared and Laser Engineering, 2018, 47(4):0404002. (in Chinese)
    [10] Hao X P, Song J, Sun J P, et al. Vacuum radiance temperature national standard facility for infrared remote sensors of Chinese Fengyun meteorological satellites[J].Optics and Precision Engineering, 2015, 23(7):1845-1851.
    [11] Shu Xin, Hao Xiaopeng, Song Jian, et al. Research of 100-400 K vacuum infrared radiance temperature standard blackbody source[J]. Acta Metrologica Sinica, 2019, 40(1):13-19. (in Chinese)
    [12] Xue Xiangqun, Li Jianjun, Zheng Xiaobing, et al. Investigation of mid-infrared radiance calibration based on parametric amplification[J]. Journal of Atmospherjc and Environmental Optics, 2010, 5(5):358-365. (in Chinese)
    [13] Jeppe Seidelin D, Christian P, Peter T L. Theory for up-conversion of incoherent images[J]. Optics Express, 2012, 20(20):1475-82.
    [14] Brida G, Genovesea M, Novero C. On the measurement of photon flux in parametric down-conversion[J]. The European Physical Journal D, 2000, 8(2):273-275.
  • [1] 丁经纬, 郝小鹏, 于坤, 宋健, 周晶晶.  黑体涂层光谱发射率特性研究 . 红外与激光工程, 2023, 52(10): 20230033-1-20230033-10. doi: 10.3788/IRLA20230033
    [2] 马秀秀, 王海燕, 韩启金, 张学文, 赵航, 徐兆鹏, 曾健, 马灵玲, 王宁.  GF5B热红外通道星上定标与验证 . 红外与激光工程, 2023, 52(4): 20220644-1-20220644-11. doi: 10.3788/IRLA20220644
    [3] 宋健, 郝小鹏, 丁雷, 李凯, 孙建平.  真空低背景红外高光谱亮温计量标准装置研制 . 红外与激光工程, 2019, 48(10): 1004001-1004001(7). doi: 10.3788/IRLA201948.1004001
    [4] 盛一成, 顿雄, 金伟其, 郭一新, 周峰, 肖思.  星上红外遥感相机的辐射定标技术发展综述 . 红外与激光工程, 2019, 48(9): 904001-0904001(13). doi: 10.3788/IRLA201948.0904001
    [5] 谢臣瑜, 李健军, 庞伟伟, 夏茂鹏, 郑小兵.  双通道热红外标准辐亮度计定标方法研究 . 红外与激光工程, 2018, 47(4): 404002-0404002(7). doi: 10.3788/IRLA201847.0404002
    [6] 许振领, 杨淼淼, 王敏, 刘连伟, 郭正红.  红外捕获跟踪设备辐射式仿真试验技术研究 . 红外与激光工程, 2016, 45(S1): 11-16. doi: 10.3788/IRLA201645.S104003
    [7] 杨红艳, 李家国, 朱利, 殷亚秋, 张永红, 雷秋良, 陈宜金.  基于历史数据的HJ-1B/IRS热红外通道定标与分析 . 红外与激光工程, 2016, 45(3): 304004-0304004(5). doi: 10.3788/IRLA201645.0304004
    [8] 衣小龙, 方伟, 李叶飞, 叶新, 王玉鹏.  太阳辐照度绝对辐射计的定标新方法 . 红外与激光工程, 2016, 45(9): 917001-0917001(7). doi: 10.3788/IRLA201645.0917001
    [9] 崔程光, 王淑荣, 李博, 王俊博, 黄煜.  紫外可见NMOS线阵图像传感器量子效率的定标研究 . 红外与激光工程, 2015, 44(9): 2844-2848.
    [10] 王超, 胡忠辉.  低温红外目标源控温技术 . 红外与激光工程, 2015, 44(3): 827-831.
    [11] 翟文超, 徐骏, 郑小兵, 丁蕾, 陆俊桦.  应用于超光谱定标的光谱可调光源光学设计 . 红外与激光工程, 2014, 43(3): 950-955.
    [12] 李斌, 吴海英, 王文涛, 刘鹏.  一种简易红外标定系统的设计 . 红外与激光工程, 2014, 43(2): 458-463.
    [13] 韩光宇, 曹立华, 张文豹.  地基目标光学辐射特性测量系统设计 . 红外与激光工程, 2014, 43(2): 551-556.
    [14] 王建军, 黄晨, 高昕, 李舰艇.  红外辐射测量系统内外标定技术 . 红外与激光工程, 2014, 43(6): 1767-1771.
    [15] 王彦斌, 王敏, 邹前进, 李华, 黄成功, 亓凤杰.  黑体模拟真实目标温度和面积的设定 . 红外与激光工程, 2014, 43(3): 779-784.
    [16] 徐骏, 孟炳寰, 翟文超, 丁蕾, 郑小兵.  基于热红外标准辐亮度计的常温黑体定标技术 . 红外与激光工程, 2014, 43(3): 716-721.
    [17] 钱婧, 孙胜利, 于胜云, 沙晟春, 盛敏健, 雍朝良.  大型红外辐射面源的设计 . 红外与激光工程, 2013, 42(1): 31-35.
    [18] 徐骏, 孟炳寰, 郑小兵, 翟文超, 李健军.  红外标准辐亮度计的研制及定标 . 红外与激光工程, 2013, 42(4): 863-868.
    [19] 易波, 张建, 屈恩世, 谢庆胜, 冷寒冰, 刘伟, 武登山, 雷杨杰, 王泽峰.  红外四谱段测温辐射计的电子学设计与实现 . 红外与激光工程, 2013, 42(10): 2630-2635.
    [20] 张伟, 谢蓄芬, 王付刚, 智喜洋.  变积分时间的空间红外相机单点绝对辐射定标法 . 红外与激光工程, 2012, 41(8): 2090-2095.
  • 加载中
计量
  • 文章访问数:  430
  • HTML全文浏览量:  78
  • PDF下载量:  25
  • 被引次数: 0
出版历程
  • 收稿日期:  2019-07-05
  • 修回日期:  2019-08-15
  • 刊出日期:  2019-12-25

基于受激参量下转换的红外标准传递辐射计定标技术

doi: 10.3788/IRLA201948.1204001
    作者简介:

    盛文阳(1991-),女,博士生,主要从事遥感器高精度定标方面的研究。Email:shengwy@mail.ustc.edu.cn

基金项目:

中国科学院实验室基金(CXJJ-17S003);国家自然科学基金(41805019)

  • 中图分类号: O432.1

摘要: 基于受激参量下转换过程产生光子的相关特性,构建了一套无需制冷且只利用可见光探测器即可实现红外标准传递辐射测量的系统。首先详细描述了该红外标准传递辐射计系统的工作原理,接着利用标准黑体拟合获取了该系统的定标系数,由此得到红外标准传递辐射计温度与待测黑体响应值的关系,最后通过与水浴黑体在相同温度点的响应值所反演获取的辐射温度作比较,进一步评估和验证了该红外标准传递辐射计系统的不确定度。实验结果表明:该红外标准传递辐射计测量水浴黑体辐亮度的联合不确定度为1.64%,表明该红外标准传递辐射计可作为标准黑体到用户传感器的传递标准。