Automatic geometric calibration method based on high frequency energy for CT

  • Received Date: 2013-01-15
  • Rev Recd Date: 2013-02-10
  • Publish Date: 2013-09-25
  • A self-calibration method based on spatial high frequency energy was proposed in view of the geometric artifacts caused by misaligned geometry in cone-beam CT. An optimization model was constructed based on spatial high frequency energy of the reconstructed image. Part of the parameters were directly extracted from the projection image to diminish the search scope, and the optimal solution of geometry parameters was achieved by NM-simplex method that makes the spatial high frequency energy of CT image maximize. To improve the speed of the algorithm, GPU was used to accelerate the process of image reconstruction to reduce the reconstruction time. Without reprocessing the reconstructed image, the proposed method has less computing complexity compared with the existing ones. The experiment results show that the presented method has significant effect in geometric artifact calibration of the reconstructed images with high accuracy and the maximum relative error is less than 5%. It can notably reduce iterations without precision reduction, and algorithm execution efficiency has been raised 18.8%.
  • [1]
    [2] Liu Yuanpeng, Zhang Fengshou. Cone-beam Computed Tomography System[M]. Beijing: Intellectual Property Publishing House, 2008. (in Chinese) 刘元朋, 张丰收. 锥束计算机断层成像系统[M]. 北京: 知识产权出版社, 2008.
    [3] Smekal L, Kachelriess M, Stepina E, et al. Geometric misalignment and calibration in cone-beam tomography[J]. Med Phys, 2004, 31(12): 3242-3266.
    [4]
    [5] Kai Yang, Alexander L, Miller F, et al. A geometric calibration method for cone beam CT systems[J]. Med Phys, 2006, 33(6): 1695-1706.
    [6]
    [7] Normand Robert, Kristina N Watt, Wang Xinying. The geometric calibration of cone-beam systems with arbitrary geometry[J]. Phys Med Biol, 2009, 54(24): 7239-7261.
    [8]
    [9]
    [10] Ford J, Zheng D, Williamson J. Estimation of CT cone-beam geometry using a novel method insensitive to phantom fabrication inaccuracy: Implications for isocenter localization accuracy[J]. Med Phys, 2011, 38(6): 2829-2840.
    [11]
    [12] Panetta D, Belcari N, Guerra A, et al. An optimization-based method for geometrical calibration in cone-beam CT without dedicated phantoms[J]. Physics in Medicine and Biology, 2008, 53(14): 3841-3861.
    [13] Patel V, Chityala R, Hoffmann K, et al. Self-calibration of a cone-beam micro-CT system[J]. Med Phys, 2009, 36(1): 48-58.
    [14]
    [15] Kyriakou Y, Lapp R, Hillebrand L, et al. Simultaneous misalignment correction for approximate circular cone-beam computed tomography[J]. Physics in Medicine and Biology, 2009, 53(22): 6267-6289.
    [16]
    [17] Kingston A, Sakallariou A, Sheppard A, et al. An auto-focus method for generating sharp 3d tomographic images[C]//Developments in X-Ray Tomography VⅡ, 2010, 7804: 78040J.
    [18]
    [19] Kinson A, Hill D, Stoyle D, et al. Automatic compensation of motion artifacts in MRI[J]. Magnetic Resonance in Medicine, 1999, 41(1): 163-170.
    [20]
    [21] Zhu Kongfeng, Jiang Wei, Wang Duanfang, et al. New kind of clarity-evaluation-function of image[J]. Infrared Laser and Engineering, 2005, 34(4): 464-468. (in Chinese) 朱孔凤, 姜威, 王端芳, 等. 一种新的图像清晰度评价函数[J]. 红外与激光工程, 2005, 34(4): 464-468.
    [22]
    [23] Kang Zongming, Zhang Li, Xie Pan. Implementation of an automatic focusing algorithm based on spatial high frequency energy and entropy[J]. Acta Electronica Sinica, 2003, 31(4): 552-555. (in Chinese) 康宗明, 张利, 谢攀. 一种基于能量和熵的自动聚焦算法[J]. 电子学报, 2003, 31(4): 552-555.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(387) PDF downloads(340) Cited by()

Related
Proportional views

Automatic geometric calibration method based on high frequency energy for CT

  • 1. Institute of Information System Engineering,Information Engineering University,Zhengzhou 450002,China;
  • 2. Hennan Province People's Hospital,Zhengzhou 450002,China

Abstract: A self-calibration method based on spatial high frequency energy was proposed in view of the geometric artifacts caused by misaligned geometry in cone-beam CT. An optimization model was constructed based on spatial high frequency energy of the reconstructed image. Part of the parameters were directly extracted from the projection image to diminish the search scope, and the optimal solution of geometry parameters was achieved by NM-simplex method that makes the spatial high frequency energy of CT image maximize. To improve the speed of the algorithm, GPU was used to accelerate the process of image reconstruction to reduce the reconstruction time. Without reprocessing the reconstructed image, the proposed method has less computing complexity compared with the existing ones. The experiment results show that the presented method has significant effect in geometric artifact calibration of the reconstructed images with high accuracy and the maximum relative error is less than 5%. It can notably reduce iterations without precision reduction, and algorithm execution efficiency has been raised 18.8%.

Reference (23)

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return