Hou Yue, Huang Kejin, Yu Guanyi, Zhang Pengquan. Development on high precision CO2 isotope measurement system based on infrared TDLAS technology[J]. Infrared and Laser Engineering, 2021, 50(4): 20200083. doi: 10.3788/IRLA20200083
Citation: Hou Yue, Huang Kejin, Yu Guanyi, Zhang Pengquan. Development on high precision CO2 isotope measurement system based on infrared TDLAS technology[J]. Infrared and Laser Engineering, 2021, 50(4): 20200083. doi: 10.3788/IRLA20200083

Development on high precision CO2 isotope measurement system based on infrared TDLAS technology

doi: 10.3788/IRLA20200083
  • Received Date: 2020-11-03
  • Rev Recd Date: 2020-12-26
  • Available Online: 2020-05-09
  • Publish Date: 2021-04-30
  • For natural gas distribution monitoring, it is very important to measure the CO2 isotope with high precision. In this paper, the tunable diode laser absorption spectrum (TDLAS) technology was adopted to realize the high precision CO2 isotope measurement through the absorption spectrum line of 13CO2/12CO2 at 4.3 μm. The measurement system consisted of a mid-infrared interband cascade laser (ICL) operating in a continuous wave mode, a long-path multipass cell (MPGC) and a mid-infrared mercury cadmium telluride (MCT) detector. Aiming at the problem that the intensity of 13CO2 and 12CO2 absorption spectra was affected by the temperature, an MPGC high precision temperature control system was developed. In the experiment, five CO2 gases of different concentrations were configured to calibrate the measurement system, and the response linearity was up to 0.999 6. The results show that when the integral time is 92 s, the isotope measurement precision is as low as 0.013 9‰, which has practical application value.
  • [1] Taylor S R. Stable isotope geochemistry [J]. Reviews of Geophysics, 2009, 17(4): 839-850.
    [2] 李春光. 基于中红外半导体激光器的气体检测系统的研制[D]. 吉林大学, 2016.

    Li Chunguang. Research and development of gas detection system based on mid-infrared semiconductor lasers[D]. Changchun: Jilin University, 2016: 13-16. (in Chinese)
    [3] Zhang Xu, Jin Weiqi, Li Li, et al. Research progress on passive infrared imaging detection technology and system performance evaluation of natural gas leakage [J]. Infrared and Laser Engineering, 2019, 48(S2): S204001. (in Chinese)
    [4] Aebersold R, Mann M. Mass spectrometry based proteomics [J]. Nature, 2003, 422(6928): 198-207. doi:  10.1038/nature01511
    [5] Chen Kai, Mei Maofei. Detection of gas concentrations based on wireless sensor and laser technology [J]. Laser Journal, 2018, 39(7): 50-54. (in Chinese)
    [6] Farajzadeh M, Mogaddam M, Aghdam S, et al. Application of elevated temperature-dispersive liquid-liquid microextraction for determination of organophosphorus pesticides residues in aqueous samples followed by gas chromatography-flame ionization detection [J]. Food Chemistry, 2016, 212: 198-204. doi:  10.1016/j.foodchem.2016.05.157
    [7] Rothman L, Gordon I, Babikov Y, et al. The HITRAN 2012 molecular spectroscopic database [J]. Journal of Quantitative Spectroscopy and Radiative Transfer, 2013, 130: 4-50. doi:  10.1016/j.jqsrt.2013.07.002
    [8] Mantz A W, Henry A, Valentin A. Stabilized tunable diode laser measurements of the P(2) line in the 13CO fundamental band broadened by helium at temperatures between 11.5 and 298.6 K [J]. Journal of Molecular Spectroscopy, 2001, 207(1): 113. doi:  10.1006/jmsp.2001.8328
    [9] Gao Wei, Cao Zhensong, Yuan Yiqian, et al. Design of a controllable low temperature cell and application [J]. Spectroscopy and Spectral Analysis, 2012, 32(3): 858-861. (in Chinese) doi:  10.3964/j.issn.1000-0593(2012)03-0858-04
    [10] Ma Hongliang, Sun Mingguo, Cao Zhensong, et al. Cryogenic cell for low-temperature spectral experiments of atmospheric molecules [J]. Optics and Precision Engineering, 2014, 22(10): 2617-2621. (in Chinese) doi:  10.3788/OPE.20142210.2617
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Figures(8)

Article Metrics

Article views(1812) PDF downloads(100) Cited by()

Related
Proportional views

Development on high precision CO2 isotope measurement system based on infrared TDLAS technology

doi: 10.3788/IRLA20200083
  • 1. College of Information Science & Technology, Beijing University of Chemical Technology, Beijing 100029, China
  • 2. School of Electronics and Information, Hangzhou Dianzi University, Hangzhou 310018, China
  • 3. School of Architecture, Tianjin University, Tianjin 300072, China

Abstract: For natural gas distribution monitoring, it is very important to measure the CO2 isotope with high precision. In this paper, the tunable diode laser absorption spectrum (TDLAS) technology was adopted to realize the high precision CO2 isotope measurement through the absorption spectrum line of 13CO2/12CO2 at 4.3 μm. The measurement system consisted of a mid-infrared interband cascade laser (ICL) operating in a continuous wave mode, a long-path multipass cell (MPGC) and a mid-infrared mercury cadmium telluride (MCT) detector. Aiming at the problem that the intensity of 13CO2 and 12CO2 absorption spectra was affected by the temperature, an MPGC high precision temperature control system was developed. In the experiment, five CO2 gases of different concentrations were configured to calibrate the measurement system, and the response linearity was up to 0.999 6. The results show that when the integral time is 92 s, the isotope measurement precision is as low as 0.013 9‰, which has practical application value.

    • 天然气与煤、石油等传统能源相比,其作为替代能源具有清洁、高效、储量丰富等特点[1-2]。为了提高天然气资源的利用率,人们在天然气资源的开发上做了大量的工作,其中最重要的是准确地确定天然气资源的分布区域。天然气形成过程中产生的稳定CO2同位素是其物理过程和化学反应的结果。因此,CO2同位素可以用来监测天然气分布[3]

      目前,监测天然气的一种非常有效的方法是检测地表溢出的气体,主要包括CO2及其碳同位素。目前高精度的碳同位素检测方法主要有质谱法[4]、色谱法[5]、火焰离子化法[6]等。但这些检测方法结构复杂,长期测量稳定性差,不适合复杂野外环境下的检测。近年来,可调谐二极管激光吸收光谱(TDLAS)技术由于具有高灵敏度和高分辨率特点,在环境、工业、生物、安全等非接触气体检测领域得到了广泛的应用。2006年,Lau教授团队在60 mbar压强和100.9 m光程条件下,采用激射中心波长为1.6 μm的分布反馈式(DFB)激光器,实现CO2同位素检测,精度达到±1.0‰。2008年,AERODYNE RESEARCH公司采用中红外吸收光谱,在76 m光程条件下,实现CO2同位素检测,精度达到0.10‰@1 s。2018年,Ghetti教授团队利用垂直空腔表面发射激光器检测从人体呼出CO2同位素,检测精度月为0.2%。虽然这些检测系统实现了CO2同位素检测,但是它们不能满足天然气探测的高性能需求,CO2同位素检测精度优于0.01‰。此外,这些检测系统复杂,不适合在野外严苛环境下使用。

      文中采用TDLAS技术,通过13CO2/12CO2在4.3 μm处的吸收谱线,实现高精度CO2同位素检测。该检测系统采用连续型中红外ICL、MPGC和高响应率的中红外碲镉汞(MCT)探测器。通过室内实验验证了该检测系统的工作性能和应用可行性。

    • 高分辨率透射分子吸收数据库(HITRAN)是计算或模拟大气分子透射和辐射的全球标准[7]。它涵盖了从微波到紫外线的广泛光谱区域。在红外波段,吸收光谱主要包括振动光谱和转动光谱,每一种气体都有多个吸收光谱。图1描述了在20 torr (1 torr =133.322 Pa )气体压强,24 m有效光程的条件下,90 ppbV CO2和1% H2O的HITRAN吸收光谱。

      Figure 1.  Absorption lines of CO2 isotope at 4.3 μm

      13CO212CO2吸收谱线的选取对于获得高精度的检测尤为重要[8]。吸收谱线对选择在2 314.36 cm−1和2 315.19 cm−1处,间距适当,不与其他化学物质重叠。该谱线对具有较强的吸收谱线强度,使检测系统具有较好的信噪比,不需要较大的长光程吸收单元,有利于检测系统的小型化。由于吸收谱线对具有相似的低能级,CO2同位素稳定性对温度的依赖性较小,特别是吸收谱线对之间没有水汽吸收谱线,检测系统不需要过度考虑无处不在的水汽影响。

    • 利用上述检测原理,检测系统组成框图如图2所示。

      Figure 2.  Block diagram of the measuring system

      检测系统主要包括光学部分和电学部分。光学部分,德国Nanoplus公司生产的具有热电制冷功能的中红外ICL作为光源。为了满足检测系统的小型化需求,MPGC的物理尺寸为20 cm×7.6 cm×10.5 cm,有效光路达到24 m。ICL的发射光以正确的位置和角度进入MPGC,在MPGC内部经过215次反射后进入中红外MCT探测器。具有热电制冷功能的中红外探测器由VIGO SYSTEM公司制造。

      在电学部分。采用自制的ICL驱动器和温度控制器代替商业仪器,减小了检测系统尺寸和成本。采用低功耗、高性能的浮点数字信号处理器(DSP)作为检测系统的控制器。在DSP的控制下,加法器将三角波信号和正弦波信号叠加起来,提供给电流源对ICL进行扫描和调制。利用自行研制的锁相放大器(LIA)对中红外MCT的信号输出进行解调。DSP利用模数转换器(ADC)获取二次谐波信号峰值,然后根据标定曲线得到13CO2/12CO2同位素。使用Bronkhorst公司生产的压力控制器(IQ+ Flow)控制MPGC在20 torr,使用KNF Neuberger公司生产的微直流泵将分离出的气体抽到MPGC中。

      由于所选两支吸收谱线的吸收系数会受到温度的影响,所以温度直接影响检测系统的精度[9-10]。文中设计了一个高精度的MPGC温度控制系统,原理框图如图3所示。

      Figure 3.  Schematic block diagram of temperature control system for MPGC

      在硬件电路方面,采用聚酰亚胺电热膜作为加热装置,高精度PT1000铂电阻作为温度传感器,构成闭环温度控制系统。软件方面采用Zieger-Nichols方法设置比例、积分、微分(P、I、D)三个参数的值。

    • 通过对MPGC工作温度进行控制,验证了研究的温度控制系统的性能,实验结果如图4所示。

      Figure 4.  Temperature control result

      图4可以看出,控制时间达到15 s时,MPGC的工作温度达到稳定状态,没有超调。在稳定状态下,温度波动小于±0.08 ℃。在达到设定的工作温度阈值之前,采用PD算法。然后,在稳定状态下使用PID算法。由于采用了上述积分分离PID控制算法,MPGC的工作温度上升较快,无超调现象。从而避免了加热超调恢复缓慢的问题。

    • 图5为5种不同浓度(20、30、40、50、60 ppmV)的二次谐波信号(1 ppm=10−6)。在检测系统响应实验中,通过减去非吸收性的背景信号,可以得到二次谐波信号。然后分别求出二次谐波信号的峰值与气体浓度之间的关系。

      Figure 5.  Second harmonic signals of five different concentrations

      图5示出,采用气体稀释系统,配置5种不同浓度的CO2气体,然后分别由本系统检测,得到不同浓度的13CO212CO2对应的二次谐波信号的峰值,进而计算CO2同位素。

    • 为了准确测量CO2同位素,有必要利用已知的气体浓度校准检测系统。将上述配置的5种不同浓度的CO2注入MPGC,测量时间为5 min。13CO212CO2的二次谐波信号峰值如图6所示(1ppb=10−9)。

      Figure 6.  Peak values of the second harmonic signal of 13CO2 and 12CO2

      图6分别测量了13CO212CO2的二次谐波信号的峰值。利用获得的实验数据,通过线性拟合得到以下公式:

      式中:max2f(12C)和max2f(13C)分别为12CO213CO2二次谐波信号的峰值。根据上述两个公式,推导出二次谐波信号的峰值,可以计算出12CO213CO2的浓度,响应线性可达0.999 6,从而可以准确得到碳同位素。

    • 在测量13CO2/12CO2同位素比值时,测量数据随时间漂移。为了测试检测系统的测量精度和长期稳定性,将一定浓度的CO2泵入MPGC。根据检测数据计算艾伦方差结果如图7所示。

      Figure 7.  Allan variances of the measured results

      实验结果表明,在积分时间为1 s的情况下,CO2同位素检测精度达到0.610‰。此外,当积分时间增加到92 s时,相应的检测精度可以显著降低至0.013 9‰。由于92 s之前白噪声是主要成分,因此检测精度随着积分时间的增加而减小。92 s之后,漂移成为主导噪音,检测精度开始上升。图中绿色虚线描述了白噪声主导下的系统响应。

    • 实验中采用两个流量控制器分别在2.5 SCCM和177.5 SCCM控制CO2气体和N2的流量,从而形成混合气体泵入MPGC。经2 h检测后,实测数据如图8所示。

      Figure 8.  CO2 isotope measurement results within two hours

      在2 h的测试实验中,CO2同位素值在−8.351‰~9.736‰之间,平均值为−9.081‰,最大波动值为0.73‰。在长期实验中,气瓶CO2气体与纯N2载气之间存在流速波动,导致了碳同位素比值的波动。

    • 文中报道了基于TDLAS技术的CO2同位素检测系统,其由中红外ICL、中红外MCT探测器和MPGC组成。CO2同位素吸收谱线分别位于2 314.36 cm−1和2 315.19 cm−1。为了实现CO2同位素的高精度检测,研制了一种高精度的MPGC温度控制系统。实验中,检测系统的响应线性度可达0.999 6%。当积分时间为92 s时,检测精度最低为0.013 9‰。最后,将该检测系统检测配置CO2气体同位素,表明其具备实际应用价值。

      在许多实际应用中,ICL受到诸如热波动和机械振动等低频因素扰动,导致激光波长略有波动。激光波长锁定技术(LWL)可以保持ICL的工作点恒定,从而减少光波长波动。因此,利用LWL技术可以进一步提高检测系统的精度,但是后续工作需要对检测系统结构重新设计。同时,可以通过增加阻尼防水装置,增强检测系统的抗干扰能力,可以进一步提高系统的稳定性。